Crystal temperatures on Oct.1-Oct.2
- Dash line denotes the start time of each run
First 6 runs taken on Oct.1-Oct.2 (run 1549-1554)
Electron energy
Energy resolution in NPS
Coefficients (zoom in)
Coefficients in 2D view (zoom in)
Last 6 runs taken on Oct.1-Oct.2 (run 1555-1560)
Electron energy
Energy resolution in NPS
Coefficients (zoom in)
Coefficients in 2D view (zoom in)
Ratio of these coefficients (last 6 runs / first 6 runs)
- The light yield of crystals decrease when the temperature increase
- Higher temperature result in higher coefficients
Energy with mis-calibration due to the unsteady temperature
- Apply the coefficients from run 1549-1554 to the data of run 1555-1560
- 32% difference of the resolution (0.025→0.033)
- Under estimation of deposited energy: 7.479→7.205 GeV ($E_{dep}=C\times Amp.$)
Electron energy
Energy resolution in NPS
Previous coefficients (with run 1437-1442) are quite uniform, why?